78 Instructions for use BiliLux
Technical data
14.10.4 Electromagnetic Immunity
The medical device is intended for use in the electromagnetic environment
specified below. The user should ensure that it is used in such an environment.
Immunity test IEC 60601 test
level
Compliance level Electromagnetic
environment
ESD
IEC 61000-4-2
± 8 kV contact
± 15 kV air
± 8 kV contact
± 15 kV air
The floors should
be wood, concrete,
or ceramic tile. If
floors are covered
with synthetic mate-
rial, the relative
humidity should be
at least 30%.
EFT
IEC 61000-4-4
± 2 kV Mains
± 1 kV I/Os
± 2 kV Mains
± 1 kV I/Os
Mains power quality
should be that of a
typical commercial
or hospital environ-
ment.
Surge
IEC 61000-4-5
± 1 kV Differential
± 2 kV Common
± 1 kV Differential
± 2 kV Common
Mains power quality
should be that of a
typical commercial
or hospital environ-
ment.
Power frequency
50/60 Hz
Magnetic field
IEC 61000-4-8
30 A/m 30 A/m Power frequency
magnetic fields
should be at levels
characteristic of a
typical location in a
typical commercial
or hospital environ-
ment.
Voltage Dips/Drop-
out
IEC 61000-4-11
100% dip for 0.5
cycle at 0 °, 45 °,
90 °, 135 °, 180 °,
225 °, 270 °, and
315 °
100% dip for 1
cycle at 0 °
60% dip for 5
cycles
30% dip for 25/30
cycles
(50 Hz/60 Hz) at 0 °
Dropout:
100% dropout
250/300 cycles
(50 Hz/60 Hz)
100% dip for 0.5
cycle at 0 °, 45 °,
90 °, 135 °, 180 °,
225 °, 270 °, and
315 °
100% dip for 1
cycle at 0 °
60% dip for 5
cycles
30% dip for 25/30
cycles
(50 Hz/60 Hz) at 0 °
Dropout:
100% dropout
250/300 cycles
(50 Hz/60 Hz)
Mains power quality
should be that of a
typical commercial
or hospital environ-
ment. If the user of
the BiliLux requires
continued operation
during power mains
interruptions, it is
recommended that
the BiliLux be pow-
ered from an unin-
terruptible power
supply or battery.

Table of Contents

Other manuals for Dräger BiliLux

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger BiliLux and is the answer not in the manual?

Dräger BiliLux Specifications

General IconGeneral
BrandDräger
ModelBiliLux
CategoryMedical Equipment
LanguageEnglish

Related product manuals