Supplement Fabius plus XL SW 3.n 11
Overview
Product labels
The following product labels have changed:
Product label Explanation
When connecting auxiliary devices, be aware of
the leakage current.
Transport information, see chapter "Preparing for
storage or transport".
Use of the external fresh-gas outlet with an addi-
tional switch:
When a controlled ventilation mode (Volume Con-
trol, Pressure Control, Pressure Support,
SIMV/PS) or the ManSpont ventilation mode is
used, set the switch on the external fresh-gas out-
let to the COSY position.
For further information, see "Using the external
fresh-gas outlet with an additional switch
(optional)" in the "Operation" chapter in the instruc-
tions for use.