Supplement Fabius plus XL SW 3.n 31
Reprocessing
Reprocessing list
Components Surface disinfec-
tion with cleaning
Manual cleaning fol-
lowed by disinfec-
tion by immersion
Machine clean-
ing with thermal
disinfection
Steam steriliza-
tion
Device surfaces Yes No No No
Breathing system
housing
Yes Yes Yes Yes
Ports No Yes Yes Yes
APL valve Yes (exterior) Yes Yes Yes
Inspiratory valve,
expiratory valve
No Yes Yes Yes
Bag elbow Yes Yes Yes Yes
Rigid arm for
breathing bag
(optional)
Yes Yes Yes Yes
Flexible arm for
breathing bag
(optional)
Yes Yes Yes Yes
Absorber con-
tainer and
absorber insert
No Yes Yes Yes
Clic adapter
(option)
Yes Yes Yes Yes
Ventilator lid Yes No Yes Yes
Ventilator mem-
brane
No No Yes Yes