EasyManuals Logo

Dräger Fabius plus XL User Manual

Dräger Fabius plus XL
44 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #31 background imageLoading...
Page #31 background image
Supplement Fabius plus XL SW 3.n 31
Reprocessing
Reprocessing list
Components Surface disinfec-
tion with cleaning
Manual cleaning fol-
lowed by disinfec-
tion by immersion
Machine clean-
ing with thermal
disinfection
Steam steriliza-
tion
Device surfaces Yes No No No
Breathing system
housing
Yes Yes Yes Yes
Ports No Yes Yes Yes
APL valve Yes (exterior) Yes Yes Yes
Inspiratory valve,
expiratory valve
No Yes Yes Yes
Bag elbow Yes Yes Yes Yes
Rigid arm for
breathing bag
(optional)
Yes Yes Yes Yes
Flexible arm for
breathing bag
(optional)
Yes Yes Yes Yes
Absorber con-
tainer and
absorber insert
No Yes Yes Yes
Clic adapter
(option)
Yes Yes Yes Yes
Ventilator lid Yes No Yes Yes
Ventilator mem-
brane
No No Yes Yes

Other manuals for Dräger Fabius plus XL

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Fabius plus XL and is the answer not in the manual?

Dräger Fabius plus XL Specifications

General IconGeneral
BrandDräger
ModelFabius plus XL
CategoryMedical Equipment
LanguageEnglish

Related product manuals