Assembly and preparation
12 Supplement Fabius plus XL SW 3.n
Assembly and preparation
Before first operation
Activating the battery
The following safety information has changed:
WARNING
Risk due to reduced power supply from the
internal battery
The charging capacity of the battery may be
reduced as a result of long-term storage.
If the battery has not been charged for 6
months, check its functional integrity. To do
this, charge the battery for 16 hours and
ventilate a test lung for at least 45 minutes in
battery operation.