EasyManuals Logo

Dräger Fabius plus XL User Manual

Dräger Fabius plus XL
44 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #35 background imageLoading...
Page #35 background image
Supplement Fabius plus XL SW 3.n 35
Maintenance
Steam sterilization
Use a steam sterilizer that meets the requirements
of the standard ISO 17665. Dräger recommends
steam sterilization with fractionated vacuum.
Prerequisite: The product has been cleaned and
disinfected.
1 Sterilize the product.
2 Check the product for visible damage and
replace if necessary.
Maintenance
Inspection
The following table has changed:
Inspections must be carried out regularly
according to the following guidelines and within the
specified intervals. Technical documentation is
available on request.
Safety checks
An additional check point has been added to "Check safety
features".
Checks Interval Responsible personnel
Inspection and safety checks
1)
1) Designation applies to the Federal Republic of Germany; corresponds to the “Recurring safety inspection” in the Federal
Republic of Austria
Every 12 months Service personnel
6 Check safety features:
Check the functional state of the optical
and acoustic alarm generators.
Check the functional state of the O
2 failure
alarm.
Check the locking device of the anesthetic
vaporizer.
Check the function of the power failure
alarm and the battery function.
Check the operation of the battery failure
alarm.
Check S-ORC functionality.

Other manuals for Dräger Fabius plus XL

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Fabius plus XL and is the answer not in the manual?

Dräger Fabius plus XL Specifications

General IconGeneral
BrandDräger
ModelFabius plus XL
CategoryMedical Equipment
LanguageEnglish

Related product manuals