Maintenance
36 Supplement Fabius plus XL SW 3.n
Maintenance
The following safety information has been added:
The following table has been amended:
The following table shows the maintenance
intervals:
NOTE
Only perform maintenance measures if there is
no patient connected to the device.
Component Interval Measure Responsible personnel
CO
2 absorber If colored violet Replace User
Water trap As needed or if
soiled
Replace User
Flow sensor As needed or if
calibration is no
longer possible.
Clean/replace User
O
2 sensor As needed, if cali-
bration is no lon-
ger possible, or if
the O
2 SENSOR
FAIL ! message is
displayed.
Replace User
Internal lithium battery Every 36 months Replace Experts
Lead-gel battery Every 3 years Replace Experts
Cylinder pressure reducer for
high-pressure cylinders
1)
1) optional
After 6 years Basic overhauling Experts
Pressure reducer for PIN
index
1)
After 6 years Replace Experts