For your safety and that of your patients
4 Supplement Fabius plus XL SW 3.n
Product-specific safety information
The following safety information has been amended:
The following safety information has been deleted:
WARNING
Risk of tipping over during transport
The medical device may tip over if handled
incorrectly. Observe the following when
transporting the medical device:
– The medical device may only be moved by
people who have the physical ability to do
so.
– To improve the maneuverability, transport
the device with 2 persons.
– When transporting over inclines, around
corners, or over thresholds (e.g., through
doors or in elevators), make sure that the
medical device does not bump against
anything.
– Remove any devices mounted to the
holding arms or the top of the device.
– Clear the writing tray and slide it
completely into the device.
– Do not pull the medical device over hoses,
cables, or other obstacles lying on the
floor.
– Do not activate the brake while the
medical device is being moved.
– Use only the handles provided to push or
pull the device.
WARNING
Risk of device failure
The device can fail if the power supply is
interrupted.
Always connect the device on an
uninterruptible power supply.