Supplement Fabius plus XL SW 3.n 37
Technical data
Technical data
General information
The following text has changed:
All patient-related volume values are normalized to
STAPD conditions.
All patient-related fresh-gas flow values are
normalized to STPD conditions.
Ambient conditions
The unit "psi" for ambient pressure has been added.
During operation
Temperature 10 to 35 °C (50 to 95 °F)
Ambient pressure 700 to 1060 cmH
2O (hPa) (10.9 to 15.9 psi)
Relative humidity 20 to 80 % (without condensation)
Height Up to 3000 m (9843 ft)
During storage and transport
Temperature -10 to 60 °C (14 to 140 °F)
Ambient pressure 700 to 1060 cmH
2O (hPa) (10.9 to 15.9 psi)
Relative humidity 10 to 90 % (without condensation)
The conditions for use when using additional devices can limit the environment of use of a system as a
whole. Vaporizers and anesthetic agents can limit the use of an anesthesia workstation with regard to its
temperature range and maximum fresh-gas flow. Therefore when using additional devices, follow the
associated instructions for use.

Other manuals for Dräger Fabius plus XL

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Fabius plus XL and is the answer not in the manual?

Dräger Fabius plus XL Specifications

General IconGeneral
BrandDräger
ModelFabius plus XL
CategoryMedical Equipment
LanguageEnglish

Related product manuals