Supplement Fabius plus XL SW 3.n 21
Configuration
Configuration
Configuration in Standby mode
Leakage test
The following text has changed:
Safety valve test
This test checks the functional integrity of the
safety valves.
The test results are displayed on the screen with
the leakage test results (B).
Troubleshooting in the event of a failed safety
valve test
The following safety information has changed:
Repeat leakage test. Do not operate the device
if the safety valve test continues to fail. Contact
DrägerService or the authorized local service
partner
Results of the compliance test
The value for the system compliance has changed:
WARNING
Risk of unexpectedly occurring overpressure
or negative pressure
A soiled or non-functioning safety valve is
not able to compensate for suddenly
occurring overpressure or negative pressure
in the breathing system.
Perform the leakage test before start-up of
the device. Observe the result of the safety
valve test.
System compliance
[mL/cmH
2O]
Displayed result
[mL/cmH2O]
6.0
Measured value and
PASSED

Other manuals for Dräger Fabius plus XL

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Fabius plus XL and is the answer not in the manual?

Dräger Fabius plus XL Specifications

General IconGeneral
BrandDräger
ModelFabius plus XL
CategoryMedical Equipment
LanguageEnglish

Related product manuals