Operation
20 Supplement Fabius plus XL SW 3.n
Ending operation
The following safety information has been added:
Preparing for storage or transport
The following safety information has changed:
WARNING
Risk of infection
Pathogenic germs can be transmitted
between patients by the circuit plug. If a
previously used Y-piece or filter is plugged
onto the circuit plug and if a reprocessed
component is attached later in the process
(e.g., during a leakage test), the new
component may become contaminated.
Only plug reprocessed components onto the
circuit plug.
WARNING
Risk of tipping over during transport
The medical device may tip over if handled
incorrectly. Observe the following when
transporting the medical device:
– The medical device may only be moved by
people who have the physical ability to do
so.
– To improve the maneuverability, transport
the device with 2 persons.
– When transporting over inclines, around
corners, or over thresholds (e.g., through
doors or in elevators), make sure that the
medical device does not bump against
anything.
– Remove any devices mounted to the
holding arms or the top of the device.
– Clear the writing tray and slide it
completely into the device.
– Do not pull the medical device over hoses,
cables, or other obstacles lying on the
floor.
– Do not activate the brake while the
medical device is being moved.
– Use only the handles provided to push or
pull the device.