Reprocessing
34 Supplement Fabius plus XL SW 3.n
Disinfectants
Use disinfectants that are nationally approved and
are suitable for the particular reprocessing
procedure.
Surface disinfection with cleaning
1 Remove soiling immediately. Use a cloth
dampened with disinfectant to remove soiling.
2 Perform surface disinfection.
3 After the product has been exposed to the
disinfectant for the specified contact time,
remove residual disinfectant.
4 Wipe with a cloth dampened with water (at
least drinking-water quality). Allow the product
to dry.
5 Check the product for visible soiling. Repeat
steps 1 to 5 if necessary.
6 Check the product for visible damage and
replace if necessary.
Manual cleaning followed by
disinfection by immersion
Manual cleaning
The cleaning agent that is used must have a pH of
between 9 and 12.
1 Prepare the cleaning solution in accordance
with the manufacturer's instructions.
2 Swirl the product backwards and forwards
several times in the solution. Make sure that
the solution reaches all surfaces and interior
spaces.
3 Rinse the product with water (at least drinking-
water quality) until residual cleaning agent is no
longer discernible.
4 Check the product for visible soiling. Repeat
steps 1 to 4 if necessary.
5 Check the product for visible damage and
replace if necessary.
Disinfection by immersion
1 Prepare the disinfectant solution in accordance
with the manufacturer's instructions.
2 Swirl the product backwards and forwards
several times in the solution. Make sure that
the solution reaches all surfaces and interior
spaces.
3 After the contact time has elapsed, rinse the
product with water (at least drinking-water
quality) until residual disinfectant is no longer
discernible.
4 Check the product for visible damage and
replace if necessary.
5 Thoroughly shake out residual water. Allow the
product to dry completely.
Machine cleaning with thermal
disinfection
Use a washer-disinfector that meets the
requirements of the standard ISO 15883. Dräger
recommends the use of a cart for anesthesia
accessories and ventilation accessories.
1 Securely position the product in the basket.
Ensure the following:
– All surfaces and interior spaces can be
flushed completely.
– The water can drain off freely.
2 Use a suitable cleaning agent.
3 Select a suitable cycle. Use demineralized
water for the final rinsing.
4 After the cycle has ended, check the product
for visible soiling. If necessary, repeat the cycle
or perform manual cleaning and disinfection by
immersion.
5 Check the product for visible damage and
replace if necessary.
6 Allow the product to dry completely.
WARNING
Risk due to penetrating liquid
Penetrating liquid may cause the following:
– Damage to the device
– Electric shock
– Device malfunctions
Ensure that no liquid penetrates the device.