Supplement Fabius plus XL SW 3.n 13
Assembly and preparation
Assembling the breathing system
The following chapters have changed:
Connecting the breathing hoses and the
filters
The data for the tidal volumes for adults and pediatric patients
have been changed:
1 Select suitable accessories for the respective
patient category.
Adults Pediatric patients Neonates
Tidal volume >700 mL 301 to 700 mL 50 to 300 mL <50 mL
Breathing bag 3L 2L 1L 0.5L
Breathing circuit Adults Pediatric Neonates
(or pediatric)
Filter Filter, HMEF, or HME Use filters with low
resistance and
compliance.