30 Supplement Fabius plus XL SW 3.n
Classification of device-specific components
Observe the instructions for use for the
components.
The following classification is a recommendation
from Dräger.
Non-critical
Control elements and device surfaces including:
Screen
–Softkeys
Rotary knob
–O
2 flush key
Flow control valves
APL valve
–O
2 sensor housing
Writing tray
Grip bar on trolley
Drawer handles
Standard rails on both sides
Clic adapter, Clic absorber
Circuit plug on the bag elbow
External FGO changeover switch and external
FGO port
Other surfaces which are frequently touched:
Side parts of the housings of the screen and of
other patient monitors
Accessory parts:
Storage trays
–Shelf
Hinged arms
Probes of compressed gas hoses
Mains plug
Gas cylinder valves
Transfer hose of the anesthetic gas
receiving system
Cables and hoses that lie on floor
–Brake
Surfaces which are touched less frequently:
Network cables and data cables
Compressed gas hoses
Pressure reducers
Gas cylinders
Drawer surfaces, outside and inside
Anesthetic gas receiving system
–Lamp
Holder for sample line
Semi-critical
Breathing system:
Breathing system housing
Inspiratory/expiratory ports, APL valve
Inspiratory valve, expiratory valve
Bag elbow
Rigid arm for breathing bag (optional)
Flexible arm for breathing bag (optional)
Breathing hoses
Absorber container and absorber insert
Soda lime dust filter (optional)
Sample line
When the sample line is fitted to the filter on
the Y-piece.
When the sample line is fitted directly to the
Y-piece and the filters are fitted on the
breathing system
Ventilator lid
Ventilator membrane
Ventilator hose
–Flow sensor
Critical
The device does not contain any components
classified as critical.