Supplement Fabius plus XL SW 3.n 29
Reprocessing
The following safety information has been deleted:
Classifications for reprocessing
Classification of medical devices
Medical devices and their components are
classified according to the way they are used and
the resulting risk.
WARNING
Risk of device failure
If the control areas located in the valve plate
are not sufficiently dried, this may
compromise the device function or may lead
to failure of the medical device.
After cleaning, the breathing system must be
sterilized with steam until it is completely dry.
Classifica-
tion
Explanation
Non-critical Components that come only into
contact with skin that is intact
Semi-critical Components that carry breathing
gas or come into contact with
mucous membranes or pathologi-
cally altered skin
Critical Components that penetrate skin or
mucous membranes or come into
contact with blood