Supplement Fabius plus XL SW 3.n 27
Reprocessing
Reprocessing
The chapter heading “Cleaning, Disinfection,and Sterilization”
has changed.
Disassembly
The following safety information has been added:
Information concerning dismounted
accessory parts and attached devices
Observe the instructions for use of the following
accessory parts:
Attached devices
Endotracheal suction
Hinged arms
Monitors
Sensors and cables
IT systems
–AGS
Workstation illumination
The following safety information has changed:
Removing the compact breathing system
The terminology for the O2 sensor has changed.
Before removing the compact breathing system,
the following hoses and cables must be removed:
Flow sensor cable
–O
2 sensor cable and O2 sensor
Pressure measurement hose
APL bypass hose
PEEP/PMAX hose
WARNING
Risk of infection
If no microbial filters are used, the breathing
system may become contaminated with
pathogens.
In this case, reprocess the breathing system
after every change of patient and insert a new
oxygen capsule.
CAUTION
Risk of damage to device and risk of injury
Reprocessing or sterilization can result in failure
of the O
2 sensor. Escaping substances can
damage the basic device and endanger the
patient and the user.
Do not reprocess or sterilize the O
2 sensor.

Other manuals for Dräger Fabius plus XL

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Fabius plus XL and is the answer not in the manual?

Dräger Fabius plus XL Specifications

General IconGeneral
BrandDräger
ModelFabius plus XL
CategoryMedical Equipment
LanguageEnglish

Related product manuals