Technical data
40 Supplement Fabius plus XL SW 3.n
The data for the O2 sensor have been added.
O2 sensor
(Electrochemical cell with perma-
nent measured value generation)
Response time (T90) Less than 16 seconds Measured values are not pressure
compensated.
Heat-up time after 5 minutes Error with
≤
3 % of the measured value
Drift sensitivity ±1 % of the measured value/ 8 h
Cross sensitivity 1 Vol% O
2 at 70 Vol% N2O and 5 Vol%
CO
2
With 4 Vol% halothane
or with 5 Vol% enflurane
or with 15 Vol% desflurane
or with 5 Vol% isoflurane
or with 10 Vol% sevoflurane
Measurement deviation due to
humidity
Max. ±0.02 % of the measured value per % relative humidity
No demonstrable quantitative effect on the gas measurement due to ambient conditions.
Maximum period of use of O
2
sensor cell
>12 months at 25 °C (77 °F), 50 % relative humidity, 50 % O2 in fresh
gas (or >5000 hours at 100 Vol% O
2)