DRAFT 18 May 2005
Jaundice Meter Instructions for Use (MU01380) Page 7 - 3
Troubleshooting
Symptom Possible Cause Action
The display is blank
when the Power switch
is in the On position.
The batteries are
exhausted.
Charge the battery (see
“Charging the Battery”
on page 4-1).
The display suddenly
goes blank during a
measurement.
The batteries are
exhausted.
Charge the battery (see
“Charging the Battery”
on page 4-1).
The Charger lamp
does not illuminate
when the Jaundice
Meter is placed on the
charger unit.
The Jaundice Meter is
not placed in the
charger unit correctly.
Place the Jaundice
Meter in the charger
unit aligned perpendic-
ular to the measuring
point with the display
facing forward.
The charger unit and
the AC adapter are not
plugged into an AC
outlet correctly.
Plug the charger unit
and the AC adapter into
an appropriate power
source correctly.
It is impossible to take
measurements.
The Ready lamp is not
illuminated.
Before taking a
measurement, ensure
that the Ready lamp is
illuminated.
The batteries are
exhausted.
Charge the battery (see
“Charging the Battery”
on page 4-1).