Reprocessing instructions | Perseus A500 45
Reprocessing
Fitting the water trap
● Attach the water trap (A).
Fitting the sample line
1. Screw the sample line (A) to the device side of the water trap (B).
2. Screw the sample line to the patient side of the filter. Ensure a careful and
patient-friendly arrangement (extensive information can be found in the
corresponding chapter in the instructions for use).
3. Ensure the correct course of the sample line.
4. Do not use adapters.