Alarms
Suppressed
alarm
limits
are
identified
by
the
symbol
»
YE
«
in
the
parameter
field.
National
and
European
standards
require
a
minimum
monitoring
with
some
alarm
functions.
These
standards
may
not
be
met
if
the
alarm
function
of
the
SpO2
monitoring
parameter
is
disabled.
Displaying
and
setting
alarm
limits
Alarms
can
be
displayed
and
set
from
all
three basic
screens
(standard,
data
and
trend
screens)
during
operation.
There
are
standard
alarm
limits
configured
for
the
ventilation
modes
which
may
be
used,
as
is,
see
"Setting
alarm
limits"
on
page
111
or
adjusted
individuaily
for
the
patient
concerned.
For
this
purpose,
the
alarm
limits
menu
can
be
selected
in
standby
via
the
soft
key
»alarm
limits«.
To
call
up
alarm
limits
during
operation:
©
Press
the
soft
key
»alarm
limits«.
Display
(example):
The
current
measured
values
and
their
alarm
limits
are
grouped
on
the
right.
Volume
The
curves
with
the
alarm
limits
represented
by
dashed
lines
are
displayed
on
the
left.
Example
etC02:
The
upper
alarm
limit
(50
mmHg)
and
lower
alarm
limit
(30
mmHg)
are
shown
alongside
the
measured
value
(38
mmHg).
A
disabled
alarm
limit
is
indicated
by
two
dashes
(-
-).
100
Instructions
for
Use
Primus
SW
2.n