EasyManua.ls Logo

Dräger REGARD-1 - SE Ex Sensitivity Range and Gas Application; SW1-A Position Warning; Applying Calibration Gas (SE Ex); Alarm Configuration Options

Dräger REGARD-1
64 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Configuration and calibration
Issue 5 - June 2005 Page 39 of 64 Pages
WARNING!
Do not change the position of switch SW1-A.
For new Dräger Ex sensors, range 3 (110 – 160 mV) should work for most gases. If,
during calibration, the gas reading does not reach the required value use a lower-
numbered sensitivity range. If the display reading is too high use a higher-numbered
sensitivity range.
2. Apply calibration gas
Use a calibration gas with concentration between 40% and 60% of the measuring
range.
Apply gas using calibration adaptor at 0.5 l/min flow rate.
Allow display reading to settle (about 2 to 3 minutes)
Adjust VR3 on the pellistor board until the voltage across TP1 and TP2 corresponds
to the concentration of the calibration gas.
Gas concentration
(as % of measuring range)
Voltage across
TP1 & TP2 (V)
40 2.08
45 2.24
50 2.40
55 2.56
60 2.72
If the device has a display, verify that the display reading corresponds to the gas
concentration.
Configuring alarms
There are three gas alarms. Each alarm can be set to be
rising or falling
latching or non-latching

Other manuals for Dräger REGARD-1

Related product manuals