EasyManua.ls Logo

Dräger REGARD-1 - Zero Calibration for SE Ex Unit; Calibrating SE Ex Sensitivity Range

Dräger REGARD-1
64 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Configuration and calibration
Issue 5 - June 2005 Page 38 of 64 Pages
Zero calibration
Connect a voltmeter across TP1 and TP2 on the main circuit board.
Make sure that the ambient air is free of measured gas and other interfering gases
which could influence the sensor.
Turn VR1 on the pellistor board until the voltage across TP1 and TP2 is
800 mV ± 20 mV. To increase the reading, turn VR1 anti-clockwise.
If the controller has a display, verify that it reads zero.
Calibrate span
1. Calibrate sensitivity range
Set the amplifier gain range using switches SW1-B, -C and -D.
Range SW1
No. Sensitivity A B C D
1
10 – 60 mV
ON OFF OFF
2
60 – 110 mV
OFF ON OFF
3
110 – 160 mV
OFF OFF ON
4
160 – 220 mV
OFF OFF OFF

Other manuals for Dräger REGARD-1

Related product manuals