PROPRIETARY AND CONFIDENTIAL DRAFT 5 Feb 04
6 - 7
Maintenance
WARNING:
Only facility-authorized personnel should perform preventive maintenance on the Resuscitaire
Radiant Warmer products. Preventive maintenance performed by unauthorized personnel could
result in personal injury or equipment damage.
NOTE:
For disposal of consumable materials, see “Disposal” on page 6-10.
Calibration
Qualified service personnel should completely check and calibrate the equipment at least annually.
However, we recommend quarterly. For details, refer to the Service Manual.
WARNING:
During calibration, inspect the secondary reflector directly under the warmer heater element for
particles. If particles are present, replace the heater element. Failure to do so could result in
patient injury or equipment damage.
Pneumatic System
All hoses should be visually checked every 6 months and replaced if damaged. Additionally, all hoses
should be replaced every 6 years.
Pneumatic System Block Diagram