PROPRIETARY AND CONFIDENTIAL DRAFT 5 Feb 04
2 - 18
Optional Features (Resuscitaire Birthing Room Warmer (WBR82) and
Resuscitaire Radiant Warmer (RW82) only)
Feature Dimension
Wall Supply Pressure 40 psi (276 kPa) to 75 psi (517 kPa)
Cylinder Pressure 3000 psi (20684 kPa) maximum
Cylinder Length (with VHA) 70 cm (27.5")
Cylinder Length (with VHA and Spacer) 76 cm (30")
Cylinder Diameter 12 cm (4.7") maximum
Patient Gas Supply Airway Pressure Limit,
Operator Adjustable
0 to 50 cm H
2
O (0 to 4.9 kPa) ± 10%
Fixed Airway Pressure Limit, Preset (81 400
73T/ MU11369)
60 cm H
2
O ± 10% (5.9 kPa) ± 10%
Primary Outlet Flow Control Range Supply
Pressure Limit
160.1 cm H
2
O (15.7 kPa) ± 20%
Primary Outlet Flow Control Range 0 LPM to 15 LPM
Auxiliary Supply Pressure Limit (w/o Auto-
Breath™ Infant Resuscitator feature)
160.1 cm H
2
O (15.7 kPa) ± 10% maximum
Auxiliary Flow Circuit Flow Range 0 LPM to 15 LPM
Fixed Airway Pressure Limit, Preset (81 400
70/ MU11360)
50 cm H
2
O (4.9 kPa) ± 20%
Fixed Airway Pressure Limit, Preset (P/N 81
400 73T/ MU11369)
160.1 cm H
2
O (15.7 kPa) ± 10%
Auxiliary Supply Pressure Limit (81 400 70/
MU11360)
40 cm H
2
O (3.9 kPa) ± 10%
RESUSCITATION (w/AutoBreath™ Infant
Resuscitator feature, PN 81 400 80T/
MU11373 and 81 400 81T/ MU11375)
Wall Supply Pressure 40 psi (276 kPa) to 75 psi (517 kPa)
Cylinder Pressure 3000 psi (20684 kPa) maximum
Cylinder Length (with VHA) 70 cm (27.5")
Cylinder Length (with VHA and Spacer) 76 cm (30")
Cylinder Diameter 12 cm (5") maximum
Patient Gas Supply Airway Pressure Limit,
Operator Adjustable
0 to 50 cm H
2
O (0 to 4.9 kPa) ± 10%
Fixed Airway Pressure Limit, Preset (81 400
80T/ MU11373)
50 cm H
2
O ± 20% (4.9 kPa)
Fixed Airway Pressure Limit, Preset (81 400
81T/ MU11375)
60 cm H
2
O ± 20% (5.9 kPa)
Primary Outlet Supply Pressure Limit 160.1 cm H
2
O (15.7 kPa) ± 20%
Primary Outlet Flow Range 0 LPM to 15 LPM
Auxiliary Supply Pressure Limit (81 400 81T/
MU11375)
160.1 cm H
2
O (15.7 kPa) ± 10% maximum

Table of Contents

Related product manuals