Operational Checkout Procedure
Perform the operational checkout procedure before each use and after disassembly for cleaning,
servicing, or maintenance. If the unit does not perform properly, do not use it and refer it to qualified
service personnel.
Controller
SHOCK HAZARD:
If the controller or other parts of the equipment fail to function properly, do not use the warmer. A
potential shock hazard exists within the warmer module. Refer service to qualified personnel.
Failure to do so could result in infant injury or equipment damage.
WARNING:
To prevent injury or damage to the warmer, employ two persons of sufficient strength to
adequately control the warmer during transport. Use the handle when moving the equipment.
WARNING:
For optimum stability, always lower the Variable Height Resuscitaire Radiant Warmer to its
lowest position prior to transport. Ensure that items placed on the shelves are properly secured.
Failure to do so could result in infant injury, personal injury or equipment damage.
1. Plug the AC line cord into the power connector on the rear panel.
SHOCK HAZARD:
Plug the power cord only into a properly grounded wall receptacle approved for hospital use and
of the correct voltage. Do not use extension cords or an AC receptacle box for this device.
Personal injury or equipment damage could occur.
2. Plug the power cord into an appropriate power source.
3. Turn ON the power switch. The self-test function initiates, and the following occurs:
a. The Apgar Timer, Baby Temperature, and Set Temperature digital displays show all eights.
b. All alarm indicators light, except for Power Fail.
c. All mode indicators light.
d. The >37°C indicator lights.
e. All 10 segments of the Heater Power indicator light.
f. The Procedural Silence indicator lights.
g. The Keypad Lock indicator lights.
h. The alarm sounds a high-pitched tone, a low-pitched tone, and then three beeps.
4. When the self-test function completes, the controller begins operating in Pre-Warm Mode.
5. Check Pre-Warm Mode:
a. The Pre-Warm indicator is on.
b. The Heater Power indicator displays 10 segments (100%) for 3 min.