PROPRIETARY AND CONFIDENTIAL DRAFT 5 Feb 04
4 - 15
c. The Heater Power indicator reduces to six segments (60%) for 12 min.
d. The Heater Power indicator reduces to three segments (30%).
6. Check Manual Mode:
a. Select Manual Mode by pressing the Mode Select key. The Manual indicator lights.
b. Press the Up arrow key until all of the Heater Power display segments light.
c. Press the Down arrow key until all of the Heater Power indicators turn off.
d. Connect the skin temperature probe to the skin temperature probe connector. The Baby
Temperature display comes on.
e. Hold the patient end of the skin temperature probe between your fingers.
NOTE:
The patient end of the skin temperature probe must be attached to something or a probe alarm may occur
indicating a difference of 0.4°C between the two thermistors in the skin temperature probe.
f. Set the Heater Power indicator to 100%. All segments light.
g. Wait 10 min.
h. After 10 min have elapsed, the Chk Patient indicator comes on, and the alert sounds one time.
i. Wait an additional 5 min. During this time, the alert sounds at 30 s intervals.
j. At the end of 5 min (15 min total), the heater shuts down, and the Heater Power indicators go off,
and an alarm sounds continuously and ramps up in volume.
k. Press the Silence/Reset key. The Chk Patient indicator and the alarm go off, and the heater
comes back on along with the Heater Power indicators.
7. Check the Keypad Lock key:
a. Press the Keypad Lock key. The Keypad Lock key lights, and the Mode Select key and the Up
and Down arrow keys are inoperable.
b. Press the Keypad Lock key. The Keypad Lock key light goes off, and the keypad is enabled.
8. Check Baby Mode:
a. Select Baby Mode by pressing the Mode Select key. The Baby indicator lights and the Set
Temperature display reads 36.5°C (97.7°F). In addition, the Baby Temperature indicator
flashes, and the alarm sounds (if the skin probe temperature and the setpoint are more than 1°C
apart).
b. Press the Silence/Reset key. The alarm goes off, and the Baby Temperature indicator remains
on.
9. Check Temperature Override Mode:
a. Press the Up arrow key to raise the set temperature to 37.0°C (98.6°F).
b. Press the >37°C key. The >37°C indicator comes on.
c. Press the Up arrow key to raise the set temperature to 38.0°C (100.4°F).
d. Press the Down arrow key to lower the set temperature to below 37.0°C (98.6°F). When the set
temperature falls below 37.0°C (98.6°F), the >37°C indicator goes off.