2 - 20
Accessories (WBR82 and RW82 only)
Alarms
Feature Dimension
Infusion pump/IV pole weight capacity 2.2 kg (4.5 lb) maximum
Monitor shelf weight limit capacity 4.5 kg (10 lb) maximum
Monitor shelf dimensions 35.6 cm (14.0") x 30.5 cm (12.0")
Xray cassette tray weight limit capacity 2.2 kg (4.5 lb) maximum
Xray cassette tray dimentions 36.8 cm (14.5”) x 27.9 cm (11.0”) x 1.9 cm (0.75”)
Instrument tray dimentions 33.0 cm (13”) x 22.8 cm x (9.0”)
Instrument tray weight limit capacity 2.2 kg (4.5 lb)
Feature Specification
High temperature Activates if the skin temperature probe is attached and the skin
temperature sensor reaches 39.0°C ±0.2°C (102.2°F). Resets at
38.5°C (101.3°F).
Check patient Activates in Manual Mode after 10 min. Remains on with audi-
ble alarm every 30 seconds for 5 min for a total of 15 min. The
heater then turns off.
Apgar timer Activates at the 1 min, 5 min, and 10 min Apgar timer intervals.
Power fail Activates when there is a loss of power. Stays activated for >10
min.
Probe fail Activates if the skin temperature probe fails (open or short).
System fail Indicates a system failure. Refer the unit to service immediately.
Baby temperature Activates if the baby temperature fluctuates 1°C above or below
the setpoint.
Electrical module, audio ramping
alarms (80 dBa maximum for infant)
Tone frequency: 1.2 KHz maximum with a three-stage sound
level: 15 seconds low, 15 seconds medium, and then high.
Blender module, pneumatic audio
alarm type
Vibrating reed, pneumatically powered.