PROPRIETARY AND CONFIDENTIAL DRAFT 5 Feb 04
2 - 1
Section 2
Introduction, Features,
and Specifications
Introduction
There are three versions of the Resuscitaire Radiant Warmer product: the Resuscitaire Birthing Room
Warmer (WBR82), the Resuscitaire Radiant Warmer (RW82), and the Resuscitaire Wall Mounted
Radiant Warmer (WMRW82).
Warmer Module
A controller provides Pre-Warm Mode, Manual Mode, or Baby Mode (servo skin temperature control) to
control the warmer. An examination light provides additional illumination of the mattress area. A warmer
head pivot allows the warmer to be pivoted to either side for x-ray procedures. Additionally, when the
warmer is pivoted, it continues to provide heat.
Cart/Bassinet (Resuscitaire Birthing Room Warmer (WBR82) and Cart only)
The bassinet is designed to provide maximum function and utility to aid in the care of the newborn. It is
detachable so you can transport the infant to the NICU, the general nursery, or another area of the
hospital. The side and front panels fold down to permit access to the infant. The mattress tilts at a 0°, 5°,
or 10° angle. Openings are provided on each side of the bassinet for the insertion of the accessory x-ray
cassette tray. A writing surface is available; when the writing surface is closed, it is not visible. Each cart
has a pass-through drawer and a locking front access drawer. The cart locks in place when used with a
warmer.
Bassinet (Resuscitaire Radiant Warmer (RW82) only)
The bassinet is designed to provide maximum function and utility to aid in the care of the newborn. You
may fold the side and front panels down to gain access to the infant. The mattress tilts at a 0°, 5°, or 10°
angle. Openings are provided on each side of the bassinet for the insertion of the accessory x-ray cassette
tray. A push-through drawer is supplied standard on every unit.
Controller
At power-up, the microprocessor within the controller performs a series of diagnostic tests to confirm the
proper operation of the system. During this time, all display and indicators light, except for Power Fail,
and a tone sounds.
When powered-up, the system initializes in Pre-Warm Mode, the controller starts the heater at 100%
power and maintains that setting for 3 min, reduces to 60% for 12 min, and then reduces the heater power
to 30%. During the low heat setting, the unit continues to provide heat with no low limit alarms enabled.
When operating the controller in Manual Mode, you can adjust the heater power from 0% to full power in
10% increments. After 10 min of operation in Manual Mode, the Check Patient alarm occurs. To silence