PROPRIETARY AND CONFIDENTIAL DRAFT 8 Jan 04
3 - 11
WARNING:
Ensure that flow resistance does not exceed 6 cm H
2
O at 5 LPM when adding attachments,
components, or subassemblies to the breathing system. Failure to do so could cause infant
injury.
WARNING:
When the AutoBreath™ Infant Resuscitator is in use, an alternate ventilation method must be
available. Failure to do so could result in infant injury.
WARNING:
Excessive air pressure could cause injury to the patient’s lungs. Always monitor the airway
pressure, and provide appropriate pressure relief during resuscitation.
WARNING:
For prolonged ventilation, use a heat and moisture exchanger or heated humidification. Failure
to do so could result in infant injury.
WARNING:
If the resuscitation module is used at altitudes significantly above sea level, ensure that all
resuscitation controls are properly calibrated. Failure to do so could result in infant injury.
Suction Precautions
WARNING:
Only qualified clinical personnel should use the suction capability provided on the Resuscitaire
Radiant Warmer. Before using the equipment, read the instructions. Failure to do so could result
in patient injury.
System Precautions
Accessory equipment connected to the analog and digital interfaces must be certified according
to the respective IEC standards (e.g. IEC950 for data processing equipment and IEC601-1 for
medical equipment). Furthermore, all configurations shall comply with the systems standard
IEC601-1-1. Any person who connects additional equipment to the signal input or output
configures a medical system, and is therefore responsible for ensuring that the system complies
with the system standard IEC601-1-1. If in doubt, consult technical service or your local
representative.