PROPRIETARY AND CONFIDENTIAL DRAFT 8 Jan 04
3 - 7
WARNING:
If the blender option is included, confirm that the oxygen/air blender control of the blended gas
supply module is correctly set prior to use. Failure to do so could result in infant injury.
WARNING:
Should any of the control knobs on the resuscitation module or the blender come loose for any
reason, do not attempt to refasten them. The calibration of these controls depends on the
position of the knob on the shaft. If knobs come loose, qualified service personnel must
calibrate the unit. Failure to do so could result in infant injury or equipment damage.
Resuscitation Precautions
WARNING:
Always monitor the airway pressure and set appropriate pressure relief during infant
resuscitation. Failure to do so could result in infant injury.
WARNING:
Only qualified clinical personnel should use the optional resuscitation module. Before using the
equipment, read the instructions. Failure to do so could result in infant injury.
WARNING:
Always operate the system with medical-grade gases. Failure to do so could result in infant
injury or equipment damage.
WARNING:
If the blender option is included, confirm that the oxygen/air blender control of the blended gas
supply module is correctly set prior to use. Failure to do so could result in infant injury.
WARNING:
Prior to use, confirm that the patient circuit contains all the parts needed. Failure to do so could
result in infant injury.
WARNING:
Ensure that the patient breathing circuit connections are secure and free of obstructions. Failure
to do so could result in infant injury.
WARNING:
The Auxiliary Outlet does not provide adjustable pressure limiting. Its gas supply pressure is
always 100% O
2
and is internally limited to 15.7 kPa (160.1 cm H
2
O). Always use a patient
resuscitation circuit that includes appropriate pressure relief. Failure to do so could result in
personal injury.
WARNING:
The auxiliary gas has a different setting that limits the maximum pressure at 3.9 kPa (40 cm
H
2
O).