PROPRIETARY AND CONFIDENTIAL DRAFT 5 Feb 04
7 - 4
Error #22 36° probe calibration error
>0.2°C
Adjust the temperature sensor of Skin Probe 1 and
Skin Probe 2.
Error #23 Ambient temperature too high
(>32°C)
Using an external thermometer, verify the ambient
temperature. If the temperature exceeds 32°C
(90°F), reduce the ambient temperature.
Error #24 There is an internal malfunc-
tion.
Refer the unit to service.
Error #25 There is an internal malfunc-
tion.
Refer the unit to service.
Error Number Explanation Corrective Action