17
Pump fault during operation. Condensate formation with cold and
damp intake air.
Use hose set with water trap (order
no. 83 21 527).
Pump outside the specified range. Have the pump checked by the
Dräger service department.
Function test has failed. Incorrect holder (pump) in position. Use holder (diffusion).
Adapter is incorrectly positioned on
the holder (diffusion).
Position adapter correctly on the
holder (diffusion) (refer to “Perform
a function test with gas” on page 7).
Flow test has failed. Flow test has not been performed. Perform flow test, repeat if
necessary.
Holder (pump) is not correctly
positioned.
Re-position holder (pump) and
check that it is correctly fitted.
Horn is too quiet. Volume is set too low. Set the volume using the Dräger
CC-Vision PC software (refer to
“Device configuration” on page 14).
Optical alarm signals not visible or
poorly visible.
Configuration or pattern incorrectly
set.
Configure the alarm signals using
the Dräger CC-Vision PC software
(refer to “Device configuration” on
page 14).
Life signal is not functioning. Configuration is set incorrectly. Configure the life signal using the
Dräger CC-Vision PC software
(refer to “Device configuration” on
page 14).
Gas detection instrument not
detected.
IR interface dirty. Clean IR interface.
Incompatible gas detection
instrument.
Use Dräger X-am 5x00.
Incorrect software version in the gas
detection instrument.
Have a software update performed
by the Dräger service department.
IR interface faulty, clip on gas
detection instrument not correctly
positioned.
Close clip on Dräger X-am 5x00.
Holder is not correctly positioned. Re-position holder and check that it
is correctly fitted.
Dräger X-am 5x00 switches off
quickly; no power supply.
Power contacts dirty or damp. Clean power contacts.
Device defect displayed. Dräger X-am 5x00 moved away
from Dräger X-zone 5000 during
operation.
Acknowledge alarm on Dräger X-
zone 5000, switch off Dräger X-zone
5000.
Fault Cause Remedy