14. Recommended Calibration Adjustment Pressures
Sensors Pressures
IRS and TRS (absolute) 20% of the full-scale range of the sensor
80% of the full-scale range of the sensor
IPS (gauge) 80% of the negative range
0 pressure
80% of the positive range.