Appendix A. PACE Menus and Screens
A.4 PACE Indicator Screens
Calibrating
Apply the lowest range pressure
(point 1 of 3)
Note the applied value
Accept when measured value is
stable
Apply the mid range pressure
(point 2 of 3)
Note the applied value
Accept when measured value is
stable
Keep the second calibration
point? (point 2 of 3)
Accept to continue
Press Retry to repeat
Measure/Enter
Keep the third calibration point?
(point 3 of 3)
Accept to continue
Press Retry to repeat
Measure/Enter
Calibration complete. accept
the calibration to return to the
calibration menu. Press
Escape to reject the
calibration.