Dynatron® 850plus & 550plus
Treatment Instructions - RUSSIAN & BIPHASIC
33
During a Biphasic or Russian treatment you may make the following modifications:
Use the toggle keys to select a different contraction/rest cycle.
Use the toggle keys to select a different ramp time.
Use the Time arrow keys to increase or decrease the treatment time.
Use the Intensity arrow keys to increase or decrease the intensity (separately for each
channel).
Use the Time toggle key to select the Pulse Rate or Width, then use the Time
up/down arrow keys to change the pulse rate (not available for Reciprocal treatments
nor when two “Normal” treatments are running simultaneously).
11. Press STOP if it is necessary to end the treatment before the treatment time has expired.
When the treatment time has elapsed, current to the patient is stopped and a tone sounds
notifying you of the treatment end. Treatments in progress may be stopped at any time
using one of the following methods.
STOP: Press this key to stop treatment at all channels.
REMOTE STOP: During “unattended therapy,” a patient stop button is provided to
allow the patient to stop treatment. The remote cord should be plugged into the machine
prior to starting the treatment. When the remote stop button is pressed, all output from
the device is stopped, time is set to zero, and a tone sounds briefly.
STOP ONE TREATMENT ONLY: If you have more than one treatment in progress,
you can stop one treatment by either of the following methods. First, press the channel
toggle key to select the channel to be stopped (that channel’s light is green when
selected).
FUNCTION-STOP. Press and hold the FUNCTION key and press STOP.
REDUCE THE TREATMENT TIME TO ZERO. Press the Time down arrow
until the Time display reaches zero. The device beeps when the time reaches
zero.
The output at the selected channel is stopped, and the device then displays the parameters
for the next treatment that remains in progress (if any).
Biphasic / Russian Default Settings
The following default settings are set by the manufacturer and are selected when you select
Biphasic or Russian. You may change these defaults to your own preferred settings. See
“Setting Defaults” in this manual.
• Normal mode
10/30 contraction/rest times
Treatment time: 10 minutes
Ramp up and down time: .5 sec.