Dynatron® 850plus & 550plus
Setting Defaults
66
High Volt Default Settings
• Continuous on time
• Treatment time: 10 minutes
• Ramp up and down time: .5 sec.
• Polarity: Negative
• Pulse Rate
Default High Range: 80-120 Hz
Default Low Range: 1-10 Hz
Available Range: 1-200 Hz
Microcurrent Default Settings
The Microcurrent treatment setup with electrodes is the default, and the following defaults are
selected:
• Bipolar wave
• .3 Hz frequency
• 50 µA intensity
• Time: 20 minutes, counts down
• Audible Tone: Off
If Probe treatment is selected, the following defaults are automatically selected:
• Negative unipolar wave
• 4 Hz frequency
• 300 µA intensity
• Time: zero, counts up
• Audible Tone: On
Available Ranges for both Electrodes and Probes:
• Frequency: Between .1 and 500 Hz
• Intensity: Between 10 and 990 microamps (in increments of 10µA).
NOTE: For Microcurrent treatments you may save separate default settings for a probe
treatment and for a treatment with electrodes. You may also indicate the most preferred
option (between probes and electrodes). The last default setting saved (probes or electrodes)
becomes the primary default when Microcurrent is selected.
Ultrasound Default Settings
• 1 MHz for 2, 5, and 10cm
2
soundheads, and 2 MHz for the 1cm
2
soundhead
• Continuous
• Time: 5 minutes
Combination Default Settings
In Combo mode you can save new default treatment time and the preferred stim modality.
The stim settings in a Combo treatment are determined by the defaults of the that modality.
Separate ultrasound default settings may be saved for the combo treatment which will not
affect ultrasound-only treatments.