Dynatron® 850plus & 550plus
General Modality Information
67
Battery Operation
Use ONLY a Dynatronics-Approved Battery
Contact Dynatronics or your Dynatronics dealer to purchase the optional battery or to obtain
specifications for an acceptable battery that may be used.* Do not substitute another battery without first
confirming with Dynatronics that the battery you are purchasing may be used with this device.
Only use a battery that CANNOT be recharged while it is in use. Disconnect the battery charger
from the AC power source before using the battery to supply power to this device.
An optional battery is available for the Dynatron 850plus and 550plus, allowing you to
deliver up to hours of battery-powered treatments wherever power may be unavailable or
unreliable. To use the optional battery, do the following:
1. Before use, ensure the battery has been adequately charged. DISCONNECT
the battery
charging cable from the battery while it is in use for treatment.
2. Plug the battery adapter into the jack labeled BATT on the back of the Dynatron 850plus
or the 550plus device. Turn the battery pack on.
4. Set up and deliver treatments as you normally do.
5. When available battery power is reduced to a certain level, the device will flash a
message BATT LO to indicate low battery power. The treatment can continue but you
will probably be unable to set up and deliver another treatment when the current
treatment has ended.
6. When the available battery power becomes too low to continue operating the device, the
BATT LO message is again displayed, the treatment intensity is ramped down, and the
machine is then shut off. Before you can continue with battery operation of the device,
you must recharge the battery. If battery power is fully depleted, the message BATT
BAD is displayed to indicate either a bad battery or to indicate the battery must be
recharged. Any treatments that were running at that time will stop.
NOTE: Use the ON/OFF switch on the battery pack
to control power to the device during
battery operation. While the battery is attached to the Dynatron 850plus or 550plus device,
the battery’s ON/OFF switch controls power to the device, overriding the device’s own