Dynatron® 850plus & 550plus
General Modality Information
68
Battery Requirements
12 volt and at least 5 amps peak current (1.5 ampere hours minimum).
Battery adapter cord
- Cigarette lighter plug on one end (to attach to the battery pack) and a barrel plug on
the other end (to fit a .325 barrel jack).
- The cord needs to be a high quality gauge wire. Radio Shack carries a high quality
cord that is recommended. Cat. No. 270-1534D. This cord comes with a 2 amp fuse.
This needs to be replaced with a 5 amp fast blow fuse.
Note: If a low-quality gauge cord is used, you can get the BATT LO error when the
battery is not low.
Battery Life
The length of time that a unit can be used with a battery pack is dependent on several factors:
The amperage of the battery pack. A larger amperage will provide longer use.
The modality used. Ultrasound requires more power than stim modalities.
The intensity of the treatments. The higher the intensity, the higher the consumption of
power.
The use of multiple treatments. The more channels used, the more power is consumed.
The amount of charge remaining on the battery.
As a general rule, the unit may be run continuously for 30 minutes to several hours depending
on these factors.