User manual SEFELEC 5x Series  58   v1.06 
 
Choose the file by pressing on its name and drag it from the Device column to the USB Key col-
umn or vice versa depending on the desired operation. A message confirming if the copy opera-
tion has been successful is displayed briefly. 
Return to the previous menu by pressing at the top left on the icon: 
< 
Proceed in the same way for the other settings files or measurement results files  
 
NOTE 
 
Settings files are saved with the extension .JSON 
 
Measurement results files are saved in .CSV format
 
They can be read by an Excel 
spreadsheet, but they cannot be modified 
from Excel and then imported into a device.
 
 
NOTE 
 
Do not disconnect the USB key during the import and export operations. 
 
On the USB key there is a "HORUS_BACKUP" directory with the following sub-directories: 
 
Go to the desired directory and open the results file in .CSV format with Excel:
   
IDN:  EATON,SEFELEC 56-S,1900607,1.1.0.22
Parameters:  Tension= 2.00E+01 V  Temps= Auto  Montée= 0,00E+00
START ISOL
Date Time VOLT AMP OHM Result
22/11/2019 15:40:38  0.0E+00 4.70E+00 4.54E+00 ERREUR Surtension
END MEG
IDN:  EATON,SEFELEC 56-S,1900607,1.1.0.22
Parameters:  Tension= 2.000E+02 V  Temps= Auto  Montée= 0,00E+00
START ISOL
Date Time VOLT AMP OHM Result
22/11/2019 15:40:49  2.0E+02 4.58E+00 4.37E+01
22/11/2019 15:40:50  2.0E+02 4.52E+00 4.43E+01
22/11/2019 15:40:51  2.0E+02 4.49E+00 4.46E+01
22/11/2019 15:40:52  2.0E+02 4.47E+00 4.47E+01
22/11/2019 15:40:53  2.0E+02 4.46E+00 4.48E+01
22/11/2019 15:40:54  2.0E+02 4.46E+00 4.49E+01
22/11/2019 15:40:55  2.0E+02 4.45E+00 4.50E+01
22/11/2019 15:40:56  2.0E+02 4.44E+00 4.51E+01
22/11/2019 15:40:57  2.0E+02 4.44E+00 4.51E+01
22/11/2019 15:40:58  2.0E+02 4.43E+00 4.51E+01
22/11/2019 15:40:58  0.0E+00 4.43E+00 4.51E+01 BON