Precautions to be observed  
 
The measured resistance values being very low (< 1 ohm), it is necessary to perform the measurement 
in using the 4 wire method to avoid the measurement lead parasitic resistance (interfaces, adapters , ...) 
It is necessary to warrant this measurement principle when connecting the measurement unit to the test 
points. 
 
If specific connections have to be done, it is necessary to use correctly rated cables (at least 5A/mm
2
) for 
the current and to connect voltage measurement cables as close as possible to the measured points. 
 
During the test time, it is advised not to move the probe to avoid breaking the current flowing which will 
produce  sparks.  This  can  modify  the  real  total  test  time  and  can  alter  the  contact  surface  of  the 
specimen.