XS user manual v0.55 - 88 -
Error messages for the Sequence mode
(message) SAFETY LOOP
The safety loop is not closed. There is no connection between the points 1-9 and 2-10 of the C5
connector of the rear panel. Make the connection, press again on the [MEASUREMENT] green push
button to trigger the measurement mode.
(message) BOARD NOT READY
The microprocessor board cannot communicate with the insulation resistance board.
You cannot perform a measurement, get in contact with our Service department (+33 1 64 11 83 48)
(message) PARAMETER ERROR
One or several parameters set for one of the sequence test is not compatible with the sequence
operation (i.e. in hipot , line TIME= MANUAL).
Check the parameters for all the test steps.
(message) SEQUENCE ERROR
Only 1 point in a multiple ground continuity
A PAUSE must be between 2 test steps
Several consequently PAUSE
(message) SETUP ERROR
In the SETUP menu, on the line INTERFACE , the option WITHOUT has to be selected.