XS series operating manual 
 
XS user manual v0.55  - 96 - 
Information on cycle timing 
Minimum cycle timing in dielectric strength test ( hipot test) 
    Measurement conditions :  
NO (second page of the setup menu) 
 
  Test time = time between change of the MES_DCH (start test) signal and change of EOT (end of 
test) signal = 980 mS 
  Measurement parameters : same as above but with hold time = 0 
     Test time = 700 mS. In this case there is no output voltage control. 
  Measurement parameters : memory number change between each measurement 
     Test time = basic minimum test time + 1.3 S 
 
Minimum cycle timing in MEGOHMMETER 
    Measurement conditions :  
NO (second page of the setup menu) 
 
  Test time = time between change of the MES_DCH (start test) signal and change of EOT (end of 
test) signal = 1.58 S 
  Measurement parameters : memory number change between each measurement 
     Test time = basic minimum test time + 1.3 S 
 
Minimum cycle timing in GROUND CONTINUITY   
    Measurement conditions :  
NO (second page of the setup menu) 
 
  Test time = time between change of the MES_DCH (start test) signal and change of EOT (end of test) 
signal = 1.19 S 
  Measurement parameters : memory number change between each measurement 
     Test time = basic minimum test time + 1.3 S