XS series operating manual 
 
XS user manual v0.55  - 23 - 
 
 
 
CAPACITANCE mode : from 1.00M to 200G (or 2 T for option 20) with an accuracy equal to the 
NORMAL mode accuracy ±100k 
Input impedance  = 10 M  ±1% 
 
Measurement threshold  
 
The megohmmeter function can operate either with one or two measurement threshold(s) between 
50k up to 200G (2T with option 20 ). 
A measurement is PASS after comparison between the displayed value and the thresholds. 
When looking for a value only higher than a limit, the LOW THRESHOLD is used to set the limit and 
the HIGH THRESHOLD is not used by setting a value equal to the maximum resistance ( basically 
200G). 
When looking for an insulation resistance higher than a limit but not too much high ( which can 
means that the specimen is not correctly connected), then the LOW THRESHOLD is set the 
minimum insulation value expected and the HIGH THRESHOLD is set to the maximum insulation 
value expected. 
 
Setting examples: 
 
PASS  ( insulation > LOW limit ) 
FAIL ( insulation < LOW limit) 
PASS ( insulation > LOW limit + < HIGH limit) 
FAIL (insulation > HIGH limit ) 
 
* threshold not active 
 
Timer  
 
The measurement voltage can stay permanently ( timer value set to 0 sec.) or during a time 
between 1 and 999 sec. adjustable by 1 sec. step. 
Rise and fall times are adjustable in time ( second ) or in voltage increase ( dv/dt) from  
0.1 to 0.9 sec. and from 1 to 999 sec. or in  V/s from 1V/s to 500V/s.