XS series operating manual 
 
XS user manual v0.55  - 54 - 
From the parameter menu, to modify a memory number : 
Press on the [MEM:x] function key 
With the UP/DOWN arrow keys, increase or decrease the memory number (from 0 to 9). The input 
parameters lines indicate the content of each memory. 
Enter the selected memory either by pressing again on the [MEM:x] function key or by pressing the 
ENTER key. 
 
 
WARNING 
 
ANY PARAMETER MODIFICATION IS AUTOMATICALLY STORED IN THE MEMORY  
 
Insulation resistance measurement  
 
Any measurement is performed by using the parameters stored in the active memory ( see MEMx 
indication in the LCD screen right hand side ) 
Connect the measurement probe and the return lead on the specimen before starting the 
measurement by pressing on the green MEASUREMENT push button. 
 
 
 
 
Fig 7.8 
 
IMPORTANT : 
 
ALWAYS MAKE SURE THAT THE HIGH VOLTAGE INDICATOR (RED LAMP) IS NOT ON WHEN 
CONNECTING OR DISCONNECTING THE SPECIMEN. 
 
In case where a test time is set to 0, the unit remains in the measurement mode permanently until the 
[DISCHARGE] red push button is pressed. 
If a test time value has been selected, every second the value of the test time counts down one unit. 
When the displayed time reaches 0 the output voltage is cut off automatically and the last measured 
value is stored on the LCD screen. According to the insulation resistance value in comparison with the 
HIGH and LOW limits, the red LED (FAIL) or the green LED (PASS) is illuminated.