XS series operating manual 
 
XS user manual v0.55  - 65 - 
Activation of the capacitive sample detection  
   
         WARNING 
         This parameter line only concerns the SXS16.  
 
 
Fig 8.11 
 
It allows the activation of the capacitive sample detection before a dielectric strength test starts. By 
default, the capacity threshold is set to 10nF.   
 
Parameter memorization  
 
XS series unit allows to store in 10 memories (numbers 0 to 9) measurement parameters (voltage, 
threshold, time , ....). From the measurement menu, to modify the memory number: 
Press on the [MEM:x] function key 
Display shows : 
 
 
Fig 8.12 
 
With  the  UP/DOWN  arrow  keys,  increase  or  decrease  the  memory  number  (from  0  to  9).  The 
parameters recall line (area 3 of the LCD screen) indicates the content of each memory. 
Enter the selected memory either by pressing again on the [MEM:x] function key or by pressing the 
ENTER key. 
 
From the parameter menu, to modify a memory number: 
Press on the [MEM:x] function key 
With the UP/DOWN arrow keys, increase or decrease the memory number (from 0 to 9). The input 
parameters lines indicate the content of each memory. 
Enter the selected memory either by pressing again on the [MEM:x] function key or by pressing the 
ENTER key. 
 
WARNING