XS series operating manual 
 
XS user manual v0.55  - 40 - 
Connection for the ground continuity measurement  
 
 
Fig 5.7 
 
  Connect the measurement probe (CO183 or CO184) on the UA, IA, UB, IB safety banana terminals 
(GREEN terminal = CURRENT, GREY terminal = VOLTAGE). 
  Connect the specimen to be tested as follows: 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
Fig 5.8 
 
WARNING : measurement of a specimen already connected to the earth. 
When measuring a ground continuity on specimen already connected to the earth, it is mandatory 
to keep the accuracy specifications, not to connect the IB and UB terminals ( label with a cross 
over the earth symbol) to the specimen side connected to the earth.