XS series operating manual
XS user manual v0.55 - 70 -
In both modes (AUTO and MANUAL) if the leakage current flowing through the specimen under test
overruns the set breakdown detection values (in -I or IMAX modes) the unit declares breakdown with
cut off of the high voltage (at the next zero crossing).
The display shows :
Fig 8.20
The memorized voltage is the breakdown voltage and the leakage current is also memorized and is
corresponding to the IMAX value.
Press on the [DISCHARGE] red push button to cancel the memorization on the LCD screen. (The HV
indicator goes out.)
Press on the [ESC] key to escape the function