63
transmitter(W)
150KHz-80MHz
p
1
5.3
⎥
⎦
⎤
⎢
⎣
⎡
=
v
d
80MHz-800MHz
P
E
d
⎥
⎦
⎤
⎢
⎣
⎡
=
1
5.3
800MHz-2.5GHz
P
E
d
⎥
⎦
⎤
⎢
⎣
⎡
=
1
7
0.01
0.1167 0.35 0.7
0.1
0.3689 1.107 2.214
1
1.1667 3.5 7
10
3.6893 11.07 22.14
100
11.6667 35 70
For transmitters rated at a maximum output power not listed above, the recommended separation distance d
in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is
the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects, and people.