EasyManuals Logo

Ecom j2kn pro User Manual

Ecom j2kn pro
54 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #17 background imageLoading...
Page #17 background image
ecom-J2KN
pro
Page 17
The position of the measured and calculated values (gas analysis)
on the display pages is free selectable (choose „Display values“ for
<F4> Hotkey). For alteration of the existing succession respectively
personal listing, proceed as follows:
-Press <F4> to activate the function.
-select the line with the cursor keys <up/down>,
-select the measured or calculated value with the cursor
keys <right/left>,
-repeat this procedure until all modifications are completed.
-Press <F4> to deactivate the function.
Position the sampling probe in the exhaust channel so that the ther-
mocouple is fully surrounded with the gas (see picture).
Perform the measurement in the core stream of the exhaust gas
channel (probe placed in the highest gas temperature area). A trend
indication for T.Gas easies the core stream search. As long as the
display shows a + symbol, the measured temperature increases, it
means the probe tip moves towards the core stream. If a - symbol is
displayed, pull the probe out of the core stream and the temperature
sinks. If no temperature change is shown for at least 3 seconds, so
the trend indication will be deleted.
Core stream search
O2
3.2
%
CO2
13.1
%
T.Gas
184
°C
T.Air
20
°C
Gas analysis
25.11.07
+
Gas stream
Probe tip
Protection bow
  
  

Other manuals for Ecom j2kn pro

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Ecom j2kn pro and is the answer not in the manual?

Ecom j2kn pro Specifications

General IconGeneral
BrandEcom
Modelj2kn pro
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals