The position of the measured and calculated values (gas analysis)
on the display pages is free selectable (choose „Display values“ for
<F4> Hotkey). For alteration of the existing succession respectively
personal listing, proceed as follows:
-Press <F4> to activate the function.
-select the line with the cursor keys <up/down>,
-select the measured or calculated value with the cursor
keys <right/left>,
-repeat this procedure until all modifications are completed.
-Press <F4> to deactivate the function.
Position the sampling probe in the exhaust channel so that the ther-
mocouple is fully surrounded with the gas (see picture).
Perform the measurement in the core stream of the exhaust gas
channel (probe placed in the highest gas temperature area). A trend
indication for T.Gas easies the core stream search. As long as the
display shows a + symbol, the measured temperature increases, it
means the probe tip moves towards the core stream. If a - symbol is
displayed, pull the probe out of the core stream and the temperature
sinks. If no temperature change is shown for at least 3 seconds, so
the trend indication will be deleted.