8. Gas measurement
8.1. Gas analysis
After the 1-minute calibration phase, the instrument switches over to
the measurement mode. The gas measurement values can be
viewed on different display pages. Use the cursor keys to scroll the
pages.
With <F1> you can switch from the measured value display to a
menu or to „Standby“ selected before (see chapter “Adjustments”).
Possible menu options are: Soot... Oil trace, Data processing, Ad-
justments, Control, Fuel type, Eff. (C), Memory -> M, Pressure. Fur-
ther you can switch with <F1> from any menu back to the measured
value display.
With <F2> you can print and store the measured values into the in-
termediate memory at the same time.
With <F3> you can switch off the CO sensor, in order to protect the
sensor against too high concentrations. The automatic disconnection
will switch off the CO sensor at approx. 4000 ppm.
With <F4> you can switch from the measured value display to a
menu or to „Standby“ selected before (see chapter “Adjustments”).
Possible menu options are: Soot... Oil trace, Data processing, Ad-
justments, Control, Fuel type, Eff. (C), Memory -> M, Display values,
Pressure. Further you can switch with <F1> from any menu back to
the measured value display.