EasyManuals Logo

Ecom j2kn pro User Manual

Ecom j2kn pro
54 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #47 background imageLoading...
Page #47 background image
ecom-J2KN
pro
Page 47
SO
2
/NO
x
filter
In the tubing leading to the CO sensor on the top of the instrument
there is a chemical filter for filtering SO
2
and NO
x
out of the flue gas.
The filter material is manganese-4-oxide granules and should be
changed once it has turned grey (colour change: pink> brown>
black> grey> white).
Probe and hose
Depending on the frequency of use, probe and hose should be regu-
larly cleaned in order to release particle deposits and to prevent early
wearing due to corrosion.
-Release the connections at the instrument and at the probe
grip to free the hose.
-Clean the hose (flow warm water in then dry respectively blow
water drops out).
Change printer paper roll
-Release the printer cover.
-If necessary, extract the paper rest out of the printer
("Adjustments"/"Paper feed"/<OK>).
-Remove the printer shaft and place the new paper roll on the
printer shaft.
-Insert the paper end in the slot (future printed side must be ahead).
-Press ("Adjustments"/"Paper feed"/<OK>) to transport ± 10 cm
paper thru the printer.
-Place the printer shaft back in the fixation.
-Insert the paper thru the cover of the printer compartment.
-Close the printer compartment while fixing the cover.

Other manuals for Ecom j2kn pro

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Ecom j2kn pro and is the answer not in the manual?

Ecom j2kn pro Specifications

General IconGeneral
BrandEcom
Modelj2kn pro
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals